ID: Aliquam sit amet sapien et
This SOP describes the process of clinical evaluation for medical devices in accordance with regulatory requirements of MDR 2107/745 Article 62 and Annex XIV.
The clinical evaluation according to MDR is defined as a systematic and planned process to continuously generate, collect, analyze and assess the clinical data pertaining to a device in order to verify the safety and performance, including clinical benefits, of the device when used as intended by the manufacturer.
The clinical evaluation takes into account the processes and outputs outlined in risk management. It ensures the level of demonstrated safety and risk-benefit ratio that is required for our medical devices both for initial certification and for continued safe use in the market.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Pellentesque habitant morbi tristique senectus et netus et malesuada fames ac turpis egestas.
The clinical evaluation shall be thorough and objective, and take into account both favorable and unfavorable data. Its depth and extent shall be proportionate and appropriate to the nature, classification, intended purpose and risks of the device in question, as well as to the manufacturer's claims in respect of the device.
The clinical evaluation is initiated for the following reasons including others, as required:
The MEDDEV 2.7/1 Rev. 4 provides the stages for clinical evaluation which are identified in the figure below.
Figure 1 - Stages of clinical evaluation according to MEDDEV 2.7/1 Rev. 4
A clinical evaluation plan is to be established and updated for each product. The clinical evaluation plan shall occur following the creation and testing of the product as defined in the SOP Design Control. In reference to the stages identified in Figure 1, the clinical evaluation plan occurs in stage 0.
The clinical evaluation plan shall contain in it at least the following as determined by Annex XIV of MDR 2017/745:
The clinical evaluation plan shall be initiated and approved prior to the approval of the clinical evaluation report.
The clinical evaluation documentation will be created by the quality team and include members of the business and product / software for review. Additionally, a separate clinical team can be used to manage the clinical evaluation process as well as other clinical processes as needed. See Section 4.2 for more information on the clinical evaluation team.
The clinical evaluation should be conducted by a suitably qualified individual or team. The requirements of the evaluators should be in line with the nature of the device and its clinical performance risks. The choice of an evaluator should be justifiable through reference to their qualifications and documented interest. Each evaluator shall have a declaration of interest.
According to MEDDEV 2.7/1 Rev. 4 Section 6.4, as a general principle, evaluators should possess knowledge of:
With respect to the particular device, evaluators should also have knowledge of:
The evaluators should have at least the following training and experience in the relevant field:
There may be circumstances where the level of evaluator expertise may be less or different which should be documented and duly justified.
External partners can be involved in both conducting the initial clinical evaluation and the post-market clinical follow-up in order to provide higher clinical expertise for the clinical evaluation team.
The clinical evaluation plan is executed to generate the clinical evaluation report. Relevant data is collected and analyzed as part of writing the Clinical Evaluation Report to evaluate clinical performance and safety.
The clinical evaluation report shall achieve the following:
The results of the clinical evaluation and the clinical evidence on which it is based shall be documented in a clinical evaluation report which shall support the assessment of the conformity of the device. The clinical evidence together with non-clinical data generated from non-clinical testing methods and other relevant documentation shall allow the manufacturer to demonstrate conformity with the general safety and performance requirements and shall be part of the technical documentation for the device in question. Both favorable and unfavorable data considered in the clinical evaluation shall be included in the technical documentation.
Additional information that should be included in the clinical evaluation report or references within the report as needed includes:
A clinical evaluation may be based on clinical data relating to a device for which equivalence to the device in question can be demonstrated. Equivalence should be claimed according to guidance found in MDCG 2020-5. Devices considered for equivalence should be CE marked in the EU or have a justification provided for why they are not.
The following technical, biological and clinical characteristics shall be taken into consideration for the demonstration of equivalence:
For software as a medical device (SAMD), also called medical device software (MDSW), a valid clinical association should be identified in accordance with MDCG 2020-1.
A valid clinical association is understood as the extent to which the MDSW’s output (e.g. concept, conclusion, calculations) based on the inputs and algorithms selected, is associated with the targeted physiological state or clinical condition. This association should be well founded or clinically accepted (e.g. existence of a scientific framework or sufficient level of evidence). The valid clinical association of a MDSW should demonstrate that it corresponds to the clinical situation, condition, indication or parameter defined in the intended purpose of the MDSW.
It should be noted that the Valid Clinical Association may not always be readily established. Thus, the clinical performance of the MDSW can serve as an additional valid clinical association from a clinical perspective for the specific intended purpose.
The valid clinical association shall include the validation of technical performance and clinical performance independently.
Following the initial clinical evaluation, a post-market clinical follow-up (PMCF) for the medical device shall be planned to ensure its continuing safety and clinical performance during the marketing phase. The PMCF serves as input to regularly update the clinical evaluation.
The quality team (or clinical team if available) is responsible to compile and implement a PMCF Plan, including at minimum:
The corrective and preventive action (CAPA) process must be initiated according to SOP Corrective and Preventive Action if new information collected as part of the PMCF activities indicates that a balanced risk-benefit ratio and the safety of the device is in question.
All PMCF activities conducted per PMCF period are documented in a Post-Market Clinical Follow-Up Report if required.
According to MEDDEV 2.7/1 Rev. 4 section 6.2.3, the clinical evaluation should be updated at least annually if the device carries significant risks or is not yet well established or every 2 to 5 years if the device is not expected to carry significant risks and is well established, a justification should be provided. Generally, Class I devices should be updated 5 years while Class IIA and above devices should be updated annually to every 2 years, as needed provided a justification.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Pellentesque habitant morbi tristique senectus et netus et malesuada fames ac turpis egestas.